Skip to main content Skip to navigation

H179181

nucleus 90 pulse power
1H 1.471 50 W
13C 3.3 us 28 W
15N 3.12 us 175 W

Operating Instructions for the Probe PH MASDVT850W6 BL1.3 X/Y/H +14N (H179181/0001) 

  1. Table of available X frequency range

 

1H -channel

X-range

Modification mode

1H ( l/4)

14N - 29Si

49.5 – 196MHz

λ/4 short circuit screw in

1H ( l/sym)

15N - 29Si

80.2 – 209.0

λ/4 short circuit screw out, λ/Csym screw in

1H ( l/2)

29Si - 65Cu

135 - 247

λ/4 short circuit screw out, λ/Csym screw out

1H ( l/2)

29Si - 31P

135 - 349

λ/4 short circuit screw out , λ/Csym screw out, with range coil

 

 

  1. Table of available X/Y combinations

 

X/Y-combination

X

Y

Modification mode

f/MHz

f/MHz

13C / 15N - 2H

(HZ16499)

213.7

86.1-130.5 (61.4 14N)

λ/- Csym27pF – mode, with 68pF shunt for 14N (on Y-channel)